The Model 1095E was designed to protect your analyzer from Acid Aerosol Mists, have user friendly operation with a menu driven touchscreen, and to decrease maintenance time. The model 1095E features a modular design for the heatsink and electronics enclosure. This allows for simple replacement of both assemblies in the case of a failure. The electrical interconnections for the modules are now terminated in pluggable connectors, eliminating the inconvenience of removing wires from terminal blocks to replace either module.
Active heat exchanger cooling is accomplished utilizing thermoelectric Peltier devices to cool the heat exchanger walls to a low temperature creating water condensation. The water separation occurs within our unique heat exchangers whose cylindrical surface is cooled to ambient, 4°C or minus 30°C depending on the cooling stage. In the 4°C impinger, condensate flows down the cold surface as a sheet, thus minimizing the surface area in contact with the gas sample allowing minimal loss of the water soluble gases. Water is removed from gas sample contact using a peristaltic pump. Wetted heat exchanger materials are Glass/Kynar and Kynar/Kynar.
The enhanced Operator Interface is now an intuitive Menu Driven Touch-Screen system. The operation of the Freezer System is automatically controlled by the Freezer System’s PLC based on timing parameters entered through the Touch Screen Operator Interface located on the front of the unit. Chiller status and alarm conditions are also displayed on this Operator Interface. To ensure a clean & dry sample is provided to the gas analyzer, our Freezer Systems include an oil-less heated head diaphragm sample pump with sample pressure control, and a water coalescing filter with integrated water slip sensor to indicate chiller overload/failure. Optionally, a liquid/water separation membrane may be included. Vacuum/pressure switches and gauges to indicate system status and sample blockages (both automatically and visually) are integrated into the system. To assist with analyzer calibration/validation, our systems can include a calibration gas selection manifold and a block & bleed manifold to support both direct and system calibration methods.
Model 1095E |
|||||||||||||||||||||||||
|
|
![]() |
![]() |
![]() |